Switch between AAI and DDD mode pacing-What is the mechanism?

Andrzej Ząbek, Barbara Małecka, Ewa Nowosielska-Ząbek, Maciej Dębski, Krzysztof Boczar, Jacek Lelakowski

Research output: Contribution to journalArticlepeer-review

1 Citation (Scopus)

Abstract

The electrocardiogram (ECG) interpretation in patients with implantable pacemaker is often a perplexing problem. The difficulty in the device evaluation increases in the presence of novel timing cycles and additional functions. Authors describe a special function frequently encountered in Medtronic dual-chamber pacemakers and implantable cardioverter-defibrillator devices called managed ventricular pacing (MVP) and demonstrate its performance in the patient with undersensing episodes in ventricular channel. Intermittent ventricular undersensing in the device with MVP feature turned on caused repetitive mode switches between AAI and DDD mode. This report shows unexceptional occurrence of tricky ECG findings in patient with Medtronic dual-chamber device.
Original languageEnglish
Article numbere12648
JournalAnnals of Noninvasive Electrocardiology
Volume24
Early online date21 Mar 2019
DOIs
Publication statusPublished - Jul 2019